ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
Calibration of a Pt resistance thermometer (Netsushin) in magnetic fields and under pressure in the range of 1.5–300 K and below 1.5 GPa is presented. With the pressure medium, olefin olygomers, Daphne 7373, the pressure is continuous at its solidification and the pressure drop from 300 to 4.2 K by 0.15–0.17 GPa is constant, irrespective of the initial clamped pressure at 300 K. The applicability of the thermometer and the medium for precise study in field and pressure at low temperature is discussed. © 1997 American Institute of Physics.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1148145
Permalink