ALBERT

All Library Books, journals and Electronic Records Telegrafenberg

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    [S.l.] : American Institute of Physics (AIP)
    Review of Scientific Instruments 67 (1996), S. 1970-1975 
    ISSN: 1089-7623
    Source: AIP Digital Archive
    Topics: Physics , Electrical Engineering, Measurement and Control Technology
    Notes: A dc method for thermopower (TP) measurements in magnetic fields up to 17 T in the temperature range from 3 K to 300 K is introduced. A special heating technique named "seesaw heating,'' a method similar to ac methods, was developed and resulted in good thermal stability, better accuracy and a faster procedure. Chromel–constantan thermocouples were utilized for measuring probes because of small and simple temperature dependence of magnetic field effect. Finding an empirical formula for all the calibration curves of TP of Chromel enable us to measure TP at any field and at any temperature. Finally, the experimental results on ErCo2 and LaRu2 are presented as test examples to demonstrate the possibilities and accuracy of our setup. © 1996 American Institute of Physics.
    Type of Medium: Electronic Resource
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...