    Woodbury, NY : American Institute of Physics (AIP)
    Applied Physics Letters 65 (1994), S. 779-780 
    ISSN: 1077-3118
    Source: AIP Digital Archive
    Topics: Physics
    Notes: A dc SQUID gradiometer with a large effective pickup area using only a single YBa2Cu3O7−x layer and step-edge Josephson junctions on SrTiO3 substrate has been prepared. The planar gradiometer consists of two loops coupled directly to a small area SQUID in the center of the structure. The main advantage in comparison to magnetometers is the possibility of direct measurements in a magnetically unshielded environment. At 77 K without bias reversal technique we obtain a white noise level of about 4.5×10−5 Φ0 Hz−1/2 or 8×10−13 T cm−1 Hz−1/2 down to 10 Hz. At 1 Hz noise increases up to 6×10−12 T cm−1 Hz−1/2.