    Fresenius' journal of analytical chemistry 361 (1998), S. 81-85 
    ISSN: 1432-1130
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Abstract A compact system for remote and non intrusive in situ analysis of fluorescent tracers using a newly developed pulsed microchip laser coupled to fiber optics was used for in situ rhodamine determinations. By using a crystal doubling in front of the microchip Nd-YAG laser, it is possible to obtain 532 nm at 5 kHz with an energy of 0.6 μJ in a 0.5 ns pulse. Using fiber optics and a passive optode, it was possible to analyze remotely the fluorescence of rhodamine with a compact detection system (monochromator and photomultiplier). Limits of detection down to 10–10–10–11 mol/L can be reached depending on the rhodamine studied. Such a laser can be directly implanted in the optode avoiding laser losses when exciting in the U.V.
    Type of Medium: Electronic Resource