    Description / Table of Contents: Summary The air-gap gas sensing electrode according to Růžička is improved in several aspects. The reliability was increased by the use of a combined glass-reference electrode. A new way of sample introduction gave a better accuracy in practice. The electrode is used to determine carbonate concentrations in the range of 5 · 10−2–5 · 10−4M with an accuracy of 2–9%. The electrode is in use in the analysis of environmental carbonate.
    Notes: Zusammenfassung Die Luftspaltelektrode nach Růžička wurde in einigen Punkten verbessert. Ihre Zuverlässigkeit wurde durch Anwendung einer Einstabmeßkette erhöht; ein neues System der Probezugabe bietet eine bessere Genauigkeit in der Praxis. Die Elektrode wurde zur Messung von Carbonatkonzentrationen im Betrieb von 5 · 10−2–5 · 10−4 M eingesetzt, wobei die Genauigkeit 2–9% betrug. Sie wird zur Umweltanalyse von Carbonat angewendet.