ISSN:
1612-1112
Schlagwort(e):
Column liquid chromatography
;
Phenolic compounds
;
Ion-pair solid phase extraction
;
On-line trace enrichment
Quelle:
Springer Online Journal Archives 1860-2000
Thema:
Chemie und Pharmazie
Notizen:
Summary Two packing materials, C18 and PLRP-S, are studied for on-line trace enrichment of phenolic compounds in water. Various precolumns of different internal diameter are also tested and the addition of an ion-pair reagent to increase retention and thus, breakthrough volumes of phenolic compounds, is studied. Best results are obtained when a PLRP-S precolumn is coupled on-line with a C18 analytical column and DAD detector. Addition of TBA considerably increases breakthrough volumes. In contrast, when a C18 precolumn is used, breakthrough volumes are lower and it is impossible to determine TCP and PCP, under the experimental conditions used, because of interference of other nonpolar compounds in the samples. The performance of the system is evaluated with river and tap water and the preconcentration of 10 ml of sample in a PLRP-S precolumn involves a linear range between 1 μg 1−1 and 20 μl−1 and limits of determination between 0.5 μg l−1 and 1 μg l−1 are obtained.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1007/BF02274612
Permalink