Work has continued with the evaluation of the Ion Mobility Spectrometer (IMS) for the analysis of gases having low parts-per-billion (10(exp -9)) water concentration. A modified PCP, inc. MMS-160 Mobility Mass Spectrometer System was used for the analysis of ultra-pure argon and nitrogen. The MMS-160 system permits the mass-identification of unique reactant and product ions observed in the reduced-water host gases. When the water is removed to low ppb levels, higher energy reactant ions are observed. In nitrogen, distinct odd- and even-numbered nitrogen cluster ion mobility peaks are observed, as well as adduct ion peaks from the trace contaminants in the gas. Argon also produces a cluster ion mobility peak and adduct ion peaks from trace components in the gas. Levels of contaminants in these ultra-pure gases can be determined from the ion mobility spectra. A calibrated source was used to provide variable known quantities of water to the pure gas supply of the IMS.