This paper discusses a new method for determining the iron content in Chinese fir leaves and roots using a SmartChem200 automatic chemical analyzer. With a pH value of 4-5, Fe 3 + was reduced to Fe 2 + by hydroxylamine hydrochloride, and Fe 2 + and 1,10-phenanthrolinemonohydrate generated an orange complex. The iron content was calculated based on the absorbance measured at a wavelength of 510 nm. The absorbance is directly proportional to the Fe 2 + content within a certain concentration range of Fe 2 +. The absorbance had a good linear relationship with the Fe 2 + content in the range of 0.01 mgL −1 to 5 mgL −1 ; the linear regression equation of the standard curve was Y=0.0632*X-0.00002, and the correlation coefficient was 1.0000. The method is convenient and rapid, and the experimental results were reliable.