ISSN:
1432-1130
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Abstract A microelectrode for glucose determination was constructed by immobilization of glucose oxidase (GOx) on a platinized platinum (Pt) by electrochemical polymerization of a solution containing GOx, pyrrole, and a substituted pyrrole, 4-(3-pyrrolyl)-4-oxobutyric acid. Due to platinization and covering with the polymerized heteropolypyrrole (hPPy) film, the electrode prepared showed high sensitivity to H2O2 at a low potential and significantly reduced the response to electroactive compounds, such as ascorbate, urate and 4-acetamidophenol. Working at 200 mV (vs. SCE) the electrode showed a linear response to glucose from 1.6 to 10 mM with a high sensitivity of 1 μA/mM, whereas the response to 1 mM ascorbate, urate, and 4-acetamidophenol was 0.53 μA, 18 nA and 4 nA, respectively, which was about 2.5%, 1.0% and 1.0% of that at a bare electrode. The stability of the electrode was tested at intervals of three or five days, and each test lasted about two hours. After 6 months examination, only 30% of its activity was lost.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/s002160051182
Permalink