Design principles and initial results are presented for a magnetic bottle electron spectrometer of high collection efficiency (2π sr) and good energy resolution. The instrument duplicates the carefully shaped axially decreasing magnetic field configuration of Kruit and Read [J. Phys. E 16, 313 (1983)], but replaces the high-field electromagnet central to the Kruit–Read design with two simple pole pieces energized by permanent magnets. The use of permanent magnets requires modifications of the transition region (between the high-field electron production zone and the low-field drift tube analyzer) to avoid axial field reversals associated with fringing fields.