Publication Date:
2018-06-05
Description:
Approximately 16 million Americans have diabetes mellitus, which can severely impair eyesight by causing cataracts, diabetic retinopathy, and glaucoma. Cataracts are 1.6 times more common in people with diabetes than in those without diabetes, and cataract extraction is the only surgical treatment. In many cases, diabetes-related ocular pathologies go undiagnosed until visual function is compromised. This ongoing pilot project seeks to study the progression of diabetes in a unique animal model by monitoring changes in the lens with a safe, sensitive, dynamic light-scattering probe. Dynamic light scattering (DLS), has the potential to diagnose cataracts at the molecular level. Recently, a new DLS fiber-optic probe was developed at the NASA Glenn Research Center at Lewis Field for noncontact, accurate, and extremely sensitive particle-sizing measurements in fluid dispersions and suspensions (ref. 1). This compact, portable, and rugged probe is free of optical alignment, offers point-and-shoot operation for various online field applications and challenging environments, and yet is extremely flexible in regards to sample container sizes, materials, and shapes. No external vibration isolation and no index matching are required. It can measure particles as small as 1 nm and as large as few micrometers in a wide concentration range from very dilute (waterlike) dispersions to very turbid (milklike) suspensions. It is safe and fast to use, since it only requires very low laser power (10 nW to 3 mW) with very short data acquisition times (2 to 10 sec).
Keywords:
Optics
Type:
Research and Technology 1999; NASA/TM-2000-209639
Format:
application/pdf
Permalink