ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
Several new goniometers were installed on the triple-axis/four-circle diffractometer, SIN-1 at PF-BL3A to improve data acquisition ability and beam collimation. The most prominent improvement is development and installation of a new two-circle/one-linear goniometer, which was designed to mount various x-ray detectors, such as an imaging plate or a multielement solid-state detector to acquire diffraction and/or fluorescence data over a wide range of scattering angles. The new goniometer and the original four-circle goniometer have a common rotation center (i.e., the four-circle goniometer center). The new goniometer can be operated independently of the four-circle goniometer for a wide range of scattering angles. The new two-circle/one-linear goniometer consists of a tandem combination of a translation stage, η, and two-circle goniometers, κ and ξ, where these axes η, κ, and ξ correspond to polar coordinates (ρ, θ, cursive-phi), respectively. Other improvements such as development of a pair of total reflection mirrors for the collimator axis are also reported. © 1995 American Institute of Physics.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1145854