This article is concerned with cylindrical transverse electric TE011 and rectangular TE102 microwave cavity resonators commonly used in electron paramagnetic resonance (EPR) spectroscopy. In the cylindrical mode geometry considered here, the sample is along the z axis of the cylinder, dielectric disks of 1/4 wavelength thickness are placed at each end wall, and the diameter of the cylinder is set at the cutoff condition for propagation of microwave energy in a cylindrical waveguide at the desired microwave frequency. The microwave magnetic field is exactly uniform along the sample in the region between the dielectric disks and the resonant frequency is independent of the length of the cylinder without limit. The rectangular TE102 geometry is analogous, but here the microwave magnetic field is exactly uniform in a plane. A uniform microwave field along a line sample is highly advantageous in EPR spectroscopy compared with the usual sinusoidal variation, and these geometries are called "uniform field" modes. Extensive theoretical analysis as well as finite element calculation of field patterns are presented. The perturbation of field patterns caused by sample insertion as functions of the overall length of the resonator and diameter of the sample is analyzed. The article is intended to provide a basis for design of practical structures in the range of 10 to 100 GHz. © 2001 American Institute of Physics.