Abstract The application of a photoionization detector in conjunction with a flow-through system is described for the determination of arsenic, antimony, selenium, tellurium, bismuth, tin, sulfide and ammonia. The analytical procedure is based upon a preliminary generation of the hydride compounds, their subsequent transfer to the gas phase and finally their detection in the photoionization detector. Linear calibration curves were obtained over three to four orders of magnitude with the following 3s-detection limits: 4 ng/ml As, 0.3 ng/ml Sb, 0.5 ng/ml Se, 3.0 ng/ml Te, 0.1 μg/ml Bi, 0.5 μg/ml Sn, 10.0 ng/ml S2−, 0.5 μg/ml NH3.