ISSN:
0006-3592
Keywords:
on-line monitoring
;
membrane inlet mass spectrometry (MIMS)
;
penicillin fermentation
;
phenoxyacetic acid
;
Chemistry
;
Biochemistry and Biotechnology
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Biology
,
Process Engineering, Biotechnology, Nutrition Technology
Notes:
A sampling system for on-line monitoring of organic compounds of low volatility in complex fermentation media uses membrane inlet mass spectrometry (MIMS). A Syringe pump draws a continuous flow of microfiltered broth from the reactor and circulates it after acidification through a membrane inlet, in which a membrane is the only interface between the sample and the high vacuum of a mass spectrometer. All operations run automatically, i.e., sampling, acidification measurement, and calibration. The on-stream acidification enables MIMS monitoring of carboxylic acids, as they must be undissociated in order to pass the hydrophobic membrane. The performance of the monitoring system was tested by measurements of standard solutions of phenoxyacetic acid (POAA, the sie chain precursor of penicillin-V) as well as on POAA during 200 h penicillin-V fermentation. During the entire fermentation POAA was monitored n low millimolar concentrations with high accuracy and fast response to step changes in POAA concentration. Tandem mass spectrometry (MS/MS) allowed direct identification of peaks in the mass spectrum of the broth that were not accounted for by POAA. These peaks were identified as SO2 and SCO. © 1994 John Wiley & Sons, Inc.
Additional Material:
7 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/bit.260440312
Permalink