Publication Date:
2011-08-16
Description:
Two designs of plastic scintillation counters with an area of 2 sq m scanned in each case by a single photomultiplier of the FEU-49 type are described. The radial dependence of their light collection at the place of the path of the detected particle does not exceed 10% while the half width of the differential distribution of the pulse amplitudes from nonfiltered cosmic radiation at sea level is 90 to 95%, and 65%, the plastic thickness being 5 cm and 10 cm, respectively. The temperature coefficient of the counter is 0.32% per 1 C.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
Cosmic Rays (NASA-TT-F-807); p 389-393
Format:
text