ISSN:
0935-6304
Schlagwort(e):
Liquid chromatogrphy
;
Temperature increase always improves system efficiency
;
Role of Ck parameter more important than reported in the literature
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Chemie und Pharmazie
Notizen:
The effect of temperature on the efficiency of column systems in adsorptive liquid chromatography is reported. The influence of temperature changes on the values of the terms Cm2 and Ck in the HETP equation, which characterizessystems involving silica gel and benzene, ethylene chloride, butyl ether or cyclo- hexanone + heptane, is discussed for 9 substances. The results of the study have led to the following conclusions: 1Temperature increase will always improve the efficiency of typical chromatographic systems.2As expected, an essential role is played by mass transfer resistance in the mobile phase occupying the pores of the packing material (parameter Cm2).3Mass transfer resistance resulting from sorption-desorption kinetics (Ck) does not play as essential a role in the systems investigated as does Cm2, but its value is significantly higher than that reported in the literature.
Zusätzliches Material:
6 Ill.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1002/jhrc.1240040107