ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
A conventional nuclear magnetic resonance (NMR) spectrometer with an original low-frequency limit of 2 MHz and equipped with an electromagnet is rebuilt to allow performing NMR experiments at resonance frequencies down to 100 kHz. The instrument is intended for accurate field-dependent spin relaxation studies. The preamplifier and the duplexer are completely replaced by new designs that provide low noise figure, large gain and fast recovery. The receiver, the transmitter, and the magnet power supply are modified to operate at low frequency (receiver and transmitter) and at low current (power supply). The performance of the instrument is demonstrated by 2H NMR experiments in the 500 kHz–2 MHz region and is compared to that of instruments based on direct current-superconducting quantum interference device and operated on room-temperature samples. © 2000 American Institute of Physics.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1150222
Permalink