We have developed a triple-pendulum vibration isolation system for suspended mirrors used in a laser interferometer for gravitational wave detection. Owing to the passive damping method using high-performance permanent magnets, the system has a very simple, compact structure. We measured the isolation ratio for vibration below 40 Hz and confirmed that the observed and calculated values are in good agreement. The stability of the system has been demonstrated by operating a Michelson laser interferometer incorporated with a mirror mounted on the vibration isolation system.