Publication Date:
2019-07-13
Description:
Standard thermoelectric data have been determined for four thermocouple combinations used below the ice point: commercial types E, K, and T and the special combination of KP vs Au-0.07 at.% Fe. Power series coefficients necessary to generate E = f(T) data for these combinations are given. In addition, extensive tests were carried out to ascertain the variability between materials from different manufacturers. Typical wires processed for low-temperature usage often have slightly thermoelectric properties (up to 1%) from those, nominally identical, wires made for high-temperature applications. Generally, type E is the most satisfactory standardized commercial combination. The Au-0.07 at.% Fe material has been examined in particular detail, because it is not yet an ISA standard material. It has excellent sensitivity in the liquid helium/liquid hydrogen temperature range and a nearly linear thermovoltage at higher temperatures.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
Symposium on Temperature; Jun 21, 1971 - Jun 24, 1971; Washington, DC
Format: