ISSN:
0886-9383
Keywords:
Numerical deconvolution
;
Peak resolution
;
Peak separation
;
Iterative deconvolution
;
Jansson's method
;
Super-resolution
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
Severely overlapped gas chromatographic peaks can be resolved using Jansson's method. Jansson's method is an iterative nonlinear deconvolution technique and consists of Van Cittert's algorithm with the relaxation factor a function of the kth estimate of the peak shape. The maximum peak amplitude of the instrument and peak non-negativity serve as constraints to improve the peak estimate. Super-resolution is achieved without significantly degrading the chromatogram signal-to-noise ratio. The method only requires a knowledge of the instrument impulse response function and maximum peak height.
Additional Material:
7 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/cem.1180010203