Publikationsdatum:
2022-05-26
Beschreibung:
This report describes a newly developed automated Winkler titration system for dissolved oxygen in seawater which is presently in
use at the Woods Hole Oceanographic Institution. This amperometric, calculated endpoint system was compared with two
different automated and one manual Winkler method during a recent cruise. The four different methods agreed to within about 0.04
ml/l. The system described here measures the titrant needed to reach the endpoint with a resolution better than 0.001 ml. The
standard deviation of replicate samples is 0.005 ml/l and the accuracy is about 0.02 ml/l. A technique to automatically acquire
conductivity ratio measurements and calculate salinity using a Guildline Autosal Salinometer is also describe.
Beschreibung:
Funding was provided by the National Science Foundation
through Grant No. OCE88-22542.
Schlagwort(e):
Dissolved oxygen analysis
;
Automated salinity analysis
;
Hydrography
Repository-Name:
Woods Hole Open Access Server
Materialart:
Technical Report
Format:
1542542 bytes
Format: