Conclusions 1. Increased accuracy in temperature measurement is one of the main ways of improving the metrological characteristics of high-temperature test machines. 2. Stabilization of temperature with an accuracy of ΔTstab≤ΔTmeas/(2–3) should be regarded as entirely satisfactory, and it is hardly necessary to improve the existing accuracy of force measurement (±1%) for the purposes under consideration.