Publication Date:
2016-07-16
Description:
A thermal dissociation cavity attenuated phase shift spectroscopy (TD-CAPS) instrument was developed for measuring total peroxy nitrates (PNs) and organic nitrates (ONs) concentrations in the clean atmosphere. This instrument is easy to operate and can be applied to continuous measurement of PNs and ONs. A continuously measurable system is convenient to perform observations, especially in remote areas. Three lines (NO 2 , PNs, and ONs lines) were used for thermal dissociation. The NO 2 line contains a quartz tube that is not heated, while the PN and ON lines contain quartz tubes that are heated at 433 K and 633 K, respectively. The concentrations of NO 2 , NO 2 + PNs, and NO 2 + PNs + ONs can be obtained from the NO 2 , PN, and ON lines, respectively. The lower limit values of the detection limit (3σ) for PNs and ONs were estimated to be 21 parts per trillion by volume with an integration time of 2 min. PNs were selectively thermally decomposed in the PNs line and formed NO 2 quantitatively. In the ONs line, both PNs and ONs were thermally decomposed to produce NO 2 quantitatively, but partial decomposition of HNO 3 at 633 K interfered with the ONs measurement. Therefore, a HNO 3 scrubber is required before the ONs line. Continuous observations were conducted with the TD-CAPS instrument in a remote area, and the instrument performed well for obtaining PNs and ONs concentrations.
Print ISSN:
0034-6748
Electronic ISSN:
1089-7623
Topics:
Electrical Engineering, Measurement and Control Technology
,
Physics