Publication Date:
2019-07-13
Description:
The California Air Resources Board appointed an Oxidant Calibration Committee for the purpose of evaluating the accuracy of the different agency calibration procedures. The committee chose UV absorption photometry as the reference method for ozone measurement. Interagency comparisons of the various iodometric methods were conducted relative to the ultraviolet standard. The tests included versions of the iodometric methods as employed by the Air Resources Board, the Los Angeles Air Pollution Control District, and the EPA. An alternative candidate reference method for ozone measurement, gas phase titration, was also included in the test series.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
Symposium on Calibration in air monitoring; Aug 05, 1975 - Aug 07, 1975; Boulder, CO