ISSN:
0049-8246
Keywords:
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Physics
Notes:
The exponential equation R0 = R · eR0τdescribes the counting losses in wavelength-dispersive x-ray spectroscopic (WDF) detectors up to very high counting rates. The correction error can be considerably lower than 1% for a true counting rate of 7 · 105s-1. The dead time constant τ has however to be determined very accurately for this. Methods of determining τ with the desired accuracy are also described. And finally, a formula for correcting for dead time losses for two detectors in series, where each detector has a different dead time, is deduced and verified by experiment. Here again, one can calculate the true counting rate with an accuracy of 1% at a true counting rate of about 7 · 105 s-1.
Additional Material:
6 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/xrs.1300120304