Publication Date:
2011-05-25
Description:
Catalytic oxidation of H 2 on platinum: a method for in situ calibration of hygrometers Atmospheric Measurement Techniques Discussions, 4, 3083-3095, 2011 Author(s): A. W. Rollins, T. D. Thornberry, R.-S. Gao, B. D. Hall, and D. W. Fahey Standard reference samples of water vapor suitable for in situ calibration of atmospheric hygrometers are not currently widespread, leading to difficulties in unifying the calibrations of these hygrometers and potentially contributing to measurement discrepancies. We describe and evaluate a system for reliably and quantitatively converting mixtures of H 2 in air to H 2 O on a heated platinum (Pt) surface, providing a compact, portable, adjustable source of water vapor. The technique is shown to be accurate and can be used to easily and predictably produce a wide range of water vapor concentrations (≈1 ppm−2 %) on demand. The result is a H 2 O standard that is suitable for in situ calibration of hygrometers, with an accuracy nearly that of the available H 2 standards (≈±2 %).
Electronic ISSN:
1867-8610
Topics:
Geosciences