ALBERT

All Library Books, journals and Electronic Records Telegrafenberg

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    facet.materialart.
    Unknown
    Institute of Electrical and Electronics Engineers (IEEE)
    Publication Date: 2014-04-02
    Description: Optoacoustic (photoacoustic) imaging based on cylindrically focused 1-D transducer arrays comes with powerful characteristics in visualizing optical contrast. Parallel reading of multiple detectors arranged around a tissue cross section enables capturing data for generating images of this plane within micro-seconds. Dedicated small animals scanners and handheld systems using 1-D cylindrically focused ultrasound transducer arrays have demonstrated real-time cross-sectional imaging and high in-plane resolution. Yet, the resolution achieved along the axis perpendicular to the focal plane, i.e., the elevation resolution, is determined by the focusing capacities of the detector and is typically lower than the in-plane resolution. Herein, we investigated whether deconvolution of the sensitivity field of the transducer could lead to tangible image improvements. We showcase the findings on experimental measurements from phantoms and animals and discuss the features and the limitations of the approach in improving resolution along the elevation dimension.
    Print ISSN: 0278-0062
    Electronic ISSN: 1558-254X
    Topics: Medicine , Technology
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
Close ⊗