    Publication Date: 2019-06-27
    Description: Frequency bands that best discriminate the Korotkov sounds at systole and at diastole from the sounds immediately preceding these events are defined. Korotkov sound data were recorded from five normotensive subjects during orthostatic stress (lower body negative pressure) and bicycle ergometry. A spectral analysis of the seven Korotkov sounds centered about the systolic and diastolic auscultatory events revealed that a maximum increase in amplitude at the systolic transition occurred in the 18-26-Hz band, while a maximum decrease in amplitude at the diastolic transition occurred in the 40-60-Hz band. These findings were remarkably consistent across subjects and test conditions. These passbands are included in the design specifications for an automatic blood pressure measuring system used in conjuction with medical experiments during NASA's Skylab program.
    Keywords: BIOTECHNOLOGY
    Type: IEEE Transactions on Biomedical Engineering; BME-21; Mar. 197