Publication Date:
2011-08-16
Description:
The sensitivity change for a commercial fused quartz Bourdon tube precision pressure gauge, due to a change in absolute pressure level, has been analytically computed and experimentally confirmed. The computed differential pressure error is 2.5% of full scale at a 100 atm absolute pressure level. The experimental method compared the fused quartz Bourdon tube gauge digital output to the results obtained from a nitrogen gas pressure system which had a high pressure, well-type mercury manometer as the differential pressure reference.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
Review of Scientific Instruments; 43; May 1972
Format: