ISSN:
1600-5775
Source:
Crystallography Journals Online : IUCR Backfile Archive 1948-2001
Topics:
Geosciences
,
Physics
Notes:
Charge-coupled device (CCD)-based X-ray detectors allow data to be collected much more quickly ( ̃10 times) than with current on-line imaging-plate systems. At the ESRF, X-ray image intensifier/CCD detector systems have been developed. These have great potential as fast read-out detectors for macromolecular and other forms of crystallography. They are relatively large sensitive X-ray detectors but have two inherent weaknesses: convex detection surfaces leading to spatial distortion and non-uniformity of intensity response, and susceptibility to small changes in magnetic fields. A large improvement has been made to the accuracy obtained by non-uniformity of response calibration and correction, using fluorescence from doped lithium borate glasses. Monochromatic macromolecular crystallography demonstration experiments with external user groups have shown that high-quality results may be obtained under real experimental conditions.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1107/S0909049596015087