ISSN:
1045-4861
Keywords:
Chemistry
;
Polymer and Materials Science
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Medicine
,
Technology
Notes:
The preconditioned stress relaxation behavior of irradiated ultra-high molecular weight polyethylene (UHMWPE), a commonly used bearing surface in total joint replacements (TJR), was modelled in uniaxial compression at 37°C. Twenty cylindrical specimens (10 mm dia. × 27 mm length) were machined from a compression molded plate of medical grade UHMWPE and γ-irradiated to a minimum dose of 2.5 Mrad. At an initial strain rate 1000%/s, five specimens were each tested at constant strains of 1, 2, 5 and 10%. The modified superposition principle provided a good agreement of the nonlinear viscoelastic (stress relaxation) behavior of UHMWPE (standard error of 0.0753 MPa). © 1994 John Wiley & Sons, Inc.
Additional Material:
7 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jab.770050408