ISSN:
0935-6304
Keywords:
Capillary gas chromatography
;
Atomic emission detector
;
Organotin compounds
;
Organolead compounds
;
Organomercury compounds
;
Grignard derivatization
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
As part of a continuing evaluation of new analytical and sample preparation techniques conducted by the US Environmental Protection Agency (EPA), the use of capillary gas chromatography with atomic emission detection (GC-AED) for the simultaneous determination of organotin, organolead, and organomercury compounds in environmental samples was investigated. Pentylmagnesium bromide was used to pentylate ionic organotin, organolead, and organomercury compounds; the pentyl derivatives were then separated by GC and determined by AED. Several important GC-AED parameters, including the type of injector inlet, carrier gas flow rate, and helium make-up gas flow rate, were optimized for the simultaneous determination of these organometallic compounds. Their minimum detectable concentrations were approximately 1.0 to 2.5 ng/mL using a 0.5-μL on-column injection. The calibration curves exhibited good linearity between 2.5 and 2500 ng/mL for organotin and organolead compounds, and between 2.5 and 10000 ng/mL for organomercury compounds.
Additional Material:
5 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240170705