ISSN:
0935-6304
Keywords:
Element-specific GC detection
;
Spectrometric detection
;
Microwave-induced plasma
;
Detector instability
;
Solvent effect on detector stability
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
Capillary gas chromatography coupled to microwave-inducedplasma detection furnishes sensitivity adequate for trace analysis when the signal is monitored at a carbon emission line. Work performed with one of these systems at high electronic gain for the carbon signal has revealed solvent band widths in excess of four minutes; these impeded quantification of peaks eluting after the solvent. This report discusses two phenomena which could be related to the unexpected intensity of this solvent response: photomultiplier tube saturation and carbon deposits in the quartz plasma discharge tube.