Gas Chromatography, instrumentation
;
Detectors with dual ionization cells, ECD, FID
;
Applications in packed and capillary GC
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
In order to avoid splitting at the column outlet for simultaneous multidetection, a new line of detector heads with the ionization cells coupled in series has been developed.The electron capture plus flame ionization detector (EC+FI-D) has been evaluated from the point of view of the sensitivity, the linear dynamic range and separation efficiency in a capillary column GC system. The performances of each section are the same as when a single ionization detector head is used.