Electronic Resource
New York, NY
:
Wiley-Blackwell
Journal of Electron Microscopy Technique
7 (1987), S. 313-317
ISSN:
0741-0581
Keywords:
Microanalysis
;
EDS
;
HREM
;
Coherence
;
Life and Medical Sciences
;
Cell & Developmental Biology
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Natural Sciences in General
Notes:
The intensity distribution in the microdiffraction pattern depends strongly on the coherence of the illuminated source. The coherence width at the specimen level is considered as a parameter to check the coherence of an electron microscope. Two application examples are illustrated. A small interphase precipitate of MgNi2 in an Ni superalloy has been identified by a combination of microdiffraction and energy dispersive spectrum (EDS) techniques, though it can not be found by conventional transmission electron microscopy (TEM) and selected area diffraction (SAD). By means of microdiffraction as well as high-resolution electron microscopy (HREM) imaging a defect structure has been determined. Such a defect makes the lattice extinction spots appear with various intensities in the selected area diffraction and microdiffraction patterns.
Additional Material:
5 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jemt.1060070410
|
Location |
Call Number |
Expected |
Availability |