Description:
Equations were developed which describe the closed loop feedback system operation of a proposed ultrasonic, dynamic, nonintrusive flowmeter whose design is based on a constant phase, voltage controlled frequency feedback concept. These equations are based on linear feedback system theory. The time constant of a low pass filter is taken into account. The equations show that the larger the open loop gain, the smaller the error due to fluctuations in the speed of sound and the smaller the effective time constant.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type: