ISSN:
1618-2650
Keywords:
Best. von Brom in organ. Verbindungen
;
Chromatographie, Gas
;
Mikrographitküvette als Br-spezifischer Detektor
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Description / Table of Contents:
Summary A micro carbon rod atomizer has been employed for specific and quantitative bromine determinations. This Br-specific gas-chromatographic detector works by help of InBr-bands and measurement at 372.7 nm. The performance of this flame band emission detector (FBD) is demonstrated by calibration curves and the detection limits of 2 examples (ethyl bromide, ethylene dibromide), the latter being 5 and 36 ng of bromine, respectively. The reproducibility has been found to be 4.3%. An example of this detector in comparison with a thermoconductivity detector (TCD) is presented.
Notes:
Zusammenfassung Eine Mikro-Graphit-Küvette wird für spezifische quantitative Brombestimmungen nach der Indium-Methode bei 372,7 nm benützt und in Verbindung mit einem Gas-Chromatographen als Br-spezifischer Detektor eingesetzt. Die Leistungsfähigkeit dieses Detektors wird an 2 Beispielen (Äthylbromid, Dibromäthan) mit Hilfe von Eichkurven, Bestimmung von Nachweisgrenzen u. ä. demonstriert. Die Nachweisgrenzen liegen bei 5 bzw. 36 ng Brom für die 2 genannten Substanzen. Die Reproduzierbarkeit der Messungen liegt bei 4,3%. Ein Beispiel für die Registrierung eines Gemisches verschiedener Substanzen, gemessen mit dem GC + Flammen-Banden-emissions-Detektor (FBD) und zum Vergleich mit einem Wärmeleitfähigkeitsdetektor (TCD), wird dargestellt.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF00573707
Permalink