Summary
This paper describes the use of certified reference materials to monitor the long-term quality of radiochemical separations. The practical limitations which determine the actual design of the quality control are discussed. The hypothesis that the high yield of the radiochemical separation will be constant with time has been checked and validated for the elements Zn, Fe, Co, Cd, Mo and to a lesser extent for W and Th using NBS SRM 1577A, BCR CRM 274 and IAEA RM A-11. This validation could not be made for the elements Cr, Au and Ag. Especially for Cr there is a serious lack of appropriate certified reference materials.
Similar content being viewed by others
References
Woitiez JRW, Iyengar GV (1988) Fresenius Z Anal Chem 332:657–661
Woittiez JRW, de la Cruz Tangonan M, Kroon K (1990) in preparation
Heydorn K (1980) Aspects of accuracy and precision in neutron activation analysis. Report RISØ-R-419
Woittiez JRW, Geusebroek M (1989) Symposium activation techniques in the life sciences. Gaithersburg USA
Iyengar GV (1976) Jülich report Jül-1308
Samsahl K (1966) Nuleonik 8:243–256
Samsahl K (1968) Analyst 93:101–106
Samsahl K et al. (1968) Anal Chem 40:181–187
Samsahl K (1972) Sci Total Environ 1:65–74
Samsahl K et al. (1973) IAEA technical Report IAEA- 157:155–162
Kingston HM et al. (1978) Anal Chem 50:2064–2070
Greenberg RR, Kingston HM (1983) Anal Chem 55:1160
Muramatsu Y, Parr RM (1985) IAEA/RL/128:IAEA Vienna
Byrne AR et al. (1987) Fresenius Z Anal Chem 326:723–729
IAEA-TECDOC-323 (1984) International atomic energy agency, Vienna
Ihnat M (1990) personal communication; Kumpulainen J, personal communication