Abstract
Diffusion potentials are mostly not as low as assumed, they also depend on the dynamics of the measuring solution; therefore, they often are mistaken for streaming potentials at the measuring electrode.
With ground connection the reference electrodes have to conduct stray currents out of the amplifier. Many reference electrodes do not correspond to that current load. A decisive improvement can be made possible by increasing the phase boundary metal/salt solution.
Difficulties occurring with measuring mediums poor in ions consist mostly in a too large resistance of reference electro des, especially at the propagation point diaphragm/measuring solution. A slit diaphragm represents the best solution.
Zusammenfassung
Diffusionspotentiale sind meist nicht so klein, wie häufig angenommen wird, sie sind auch von der Dynamik der Meßlösung abhängig und werden deshalb oft mit Strömungspotentialen an der Meßelektrode verwechselt.
Bei Erdverbindung ergibt sich für die Bezugselektrode die Aufgabe, Störströme aus dem Verstärker abzuleiten. Dieser Strombelastung sind viele Bezugselektroden nicht gewachsen, eine entscheidende Verbesserung ist durch Vergrößern der Phasengrenzfläche Metall/Salzlösung möglich.
Die Schwierigkeiten, die bei Messungen in ionenarmen Medien auftreten, bestehen meist im zu großen Widerstand der Bezugselektrode, insbesondere an der Ausbreitungsstelle Diaphragma/Meßlösung. Spaltdiaphragmen stellen die günstigste Lösung dar.
Similar content being viewed by others
Literatur
Henderson, P.: Physik. Chem.59, 118–127 (1907).
Schwabe, K.: Wiss. Ann.4, 65–128 (1955).
Süss, R.: Arch. Techn. Mess.1960, 113–116.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Galster, H. Störungsquellen in der Bezugselektrode bei pH-Messungen. Z. Anal. Chem. 245, 62–67 (1969). https://doi.org/10.1007/BF01154252
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF01154252