Summary
The relative X-ray fluorescence (XRF) intensities for annular source and various annular specimen are measured with a Ge-Li detection system and calculated by the Monte Carlo numerical techniques (MCNT).Kα X-ray intensities of annular specimens with different radii prepared from a pure-molybdenum foil are measured and it is seen that the measured values are in good agreement with the values calculated by MCNT. The variation of the fluorescence intensity due to the collimator radius is also studied. The differences between the MCNT and experimental results are discussed in terms of the possible sources of errors.
Similar content being viewed by others
References
Bertin E.,Principle and Practice of X-Ray Spectrometric Analysis (Plenum Press, New York, N.Y.), 1975, second edition.
Jenkins R., Gould R. W. andGodeke D.,Quantitative X-Ray Spectrometry (Marcel Dekker, New York), 1981.
Weber A.,X-Ray Spectrom,12 (1983) 11.
Sparks C. J.,Adv. X-Ray Anal.,19 (1976) 19.
Giauque R. D. andJaklevic J. M.,Adv. X-Ray Anal.,15 (1971) 164.
Arinç F. andGardner R. P.,J. Radioanal. Chem.,54 (1979) 221.
de Boer D. K. G.,X-Ray Spectrom,18 (1989) 119.
Zararsiz A. andAygün E.,J. Radioanal. Nucl. Chem.,129, (1989), 367.
Tang S. M., Kump P., Yap T. andBilal M. G.,X-Ray Spectrom,15 (1986) 289.
Yap C. T. andSaligan P. P.,Nucl. Instrum. Methods A,251 (1986) 140.
Campbell J. L. andMcNelles L. A.,Nucl. Istrum. Methods,117 (1974) 519.
Storm E. andIsrael H.,Nucl. Data Tables A,7 (1970) 569.
Sartori R. andFernandez J. E.,X-Ray Spectrom.,21 (1992) 299.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Şahin, Y., Budak, G. & Karabulut, A. X-ray fluorescence intensity distribution in circular specimens. Il Nuovo Cimento D 17, 221–228 (1995). https://doi.org/10.1007/BF02451664
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02451664