Skip to main content
Log in

The Calibration and Validation of a Phase-Modulated Near-Infrared Cerebral Oximeter

  • Published:
Journal of Clinical Monitoring and Computing Aims and scope Submit manuscript

Abstract

Objective. This study was undertaken to compare the cerebral oxygenation measured by an experimental phase-modulated near-infrared (NIR) spectroscopy system with capillary saturation estimated from jugular venous oxygen saturation. Methods. Jugular venous catheters were placed in 30 patients undergoing carotid endarterectomy and 194 measurements of venous oxygen saturation were obtained intra operatively. Simultaneous measurement of optical path length at 754, 785, and 816 nm was performed using a phase-modulated near-infrared spectroscopy system. Optical calibration was performed using both an optical bench and a scattering mold. Hemoglobin saturation was calculated from NIR measurements using equations derived from diffusion theory. Capillary saturation was calculated from the arterial and venous saturations. Results. Jugular venous saturations ranged from 41 to 92%. When calibrated using the optical bench, the NIR estimates of hemoglobin saturation deviated from estimated capillary values by an average of 2.6% bias and 4.3% deviation. No systematic bias was noted. NIR values derived from mold calibration were less accurate and precise (4.6% bias and 6.9% deviation.) Use of the initial venous sample as an in vivocalibration improved the accuracy of the mold calibration but did not alter the performance of the bench calibration. Conclusions. Under the conditions tested, an experimental phase-modulated near-infrared spectroscopy system calibrated using an optical bench agreed with capillary saturation estimated from jugular venous samples. Further work is necessary to demonstrate valid performance of the system under other conditions.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

REFERENCES

  1. Levy WJ, Levin S, Chance B. Near-infrared measure-ment of cerebral oxygenation. Anesthesiology 1995; 83: 738 746

    Google Scholar 

  2. Bland JM, Altman DG. Statistical methods for assessing Levy et al: The Calibration andValidation of a Phase-ModulatedNear-Infrared CerebralOximeter 107 agreement between two methods of clinical measure-ment. Lancet 1996; 307 310

  3. Sevick EM, Chance B, Leigh J, Nioka S, Maris M. Quantitation of time-and frequency-resolved optical spectra for the determination of tissue oxygenation. Anal Biochem 1991; 95: 330 351

    Google Scholar 

  4. Wilson EM, Halsey JH. Bilateral jugular venous blood £ow by thermal dilution. Stroke 1970; 1: 348 355

    Google Scholar 

  5. Larson CP, Ehrenfeld WK, Wade JG, Wylie EJ. Jugular venous oxygen saturation as an index of adequacy of cerebral oxygenation. Surgery 1967; 62: 31 39

    Google Scholar 

  6. Kurth CD, Steven JM, Nicholson SC, Chance B, Delivoria-Papadopoulos M. Kinetics of cerebral deoxy-genation during deep hypothermic circulatory arrest in neonates. Anesthesiology 1990; 77: 656 661

    Google Scholar 

  7. DeBlasi RA, Almenrader N, Ferrari M. Brain oxygen-ation monitoring during cardiopulmonary bypass by near infrared spectroscopy. Optical Imaging of Brain Function and Metabolism II. New York: Plenum Press, 1997; 97 104

    Google Scholar 

  8. Pollard V, Prough DS, DeMelo AE, Deyo DJ, Uchida T, Widman R. The in£uence of carbon dioxide and body position on near-infrared spectroscopic assessment of cerebral hemoglobin oxygen saturation. Anesth Analg 1996; 82: 278 287

    Google Scholar 

  9. McCormick PW, Stewart M, Goetting MG, Balakrish-nan G. Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans. Stroke 1991; 22: 596 602

    Google Scholar 

  10. Mchedlishvilli GI. Arterial behavior and blood circula-tion in the brain. Plenum Press New York: Consulants Bureau, 1986; 56 57

    Google Scholar 

  11. Levy WJ. The in£uence of demographic factors on phase-modulated spectroscopy in adults. Phil Trans R Soc Lond B 1997; 352: 751 754

    Google Scholar 

  12. Harris DNF, Bailey SM, Cowans F, Wertheim D. E¡ect of optode separation on brain penetration in adults. Adv ExpMed Biol 1966; 388: 133 135

    Google Scholar 

  13. Lam JMK, Smielewski P, Al-Rawi P, Gri¤ths P, Pickard JD, Kirkpatrick PJ. Internal and external carotid contri-butions to near-infrared spectroscopy during carotid endarterectomy. Stroke 1997; 28: 906 911

    Google Scholar 

  14. Grubhofer G, Lassnigg A, Manlik F, Marx E, Trubel W, Hiesmayr M. The contribution of extracranial blood oxygenation on near-infrared spectroscopy during carotid thrombendarterectomy. Anaesthesia 1997; 52: 116 120

    Google Scholar 

  15. Delpy DT, Cope M. Quanti¢cation in tissue near-infrared spectroscopy. Phil Trans R Soc Lond B 1997; 352: 649 659

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Levy, W.J., Carpenter, J., Fairman, R.M. et al. The Calibration and Validation of a Phase-Modulated Near-Infrared Cerebral Oximeter. J Clin Monit Comput 15, 103–108 (1999). https://doi.org/10.1023/A:1009984204752

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1023/A:1009984204752

Navigation