Summary
A new absolute condensation nucleus counter was constructed to investigate the accuracy of the calibration widely used with the Pollak-Nolan photoelectric condensation nucleus counter. An attempt was made to standardize the conditions affecting calibration and, using the new counter, to recalibrate the photoelectric counter at values of extinction less than 16%.
Over this extinction range, the present work indicates nucleus concentrations which are higher by about 30% than those given by the calibration ofPollak andMetnieks [8]3). In view of the great natural variability of Aitken nucleus concentrations, this discrepancy, if real, is of little meteorological importance.
Similar content being viewed by others
References
H. Lamb,The Dynamical Theory of Sound, 2nd Edition (Dover Publications, Inc., New York 1960), 264–269.
G. McGreevy,A condensation method particle size spectrometer, Archiv. für Meteor., Geophysik und Bioklimatologie [A]14, 2 (1964).
G. McGreevy,Anomalies in measurements made with a Pollak-Nolan photoelectric nucleus counter, Archiv. für Meteor. und Bioklimatologie [A]15, 1 (1966), 90–108.
A. L. Metnieks andL. W. Pollak,Instruction for use of photoelectric condensation nucleus counters, Geophysical Bull. # 16, Dublin Inst. for Advanced Studies, School of Cosmic Physics (1959).
L. W. Pollak andJ. Daly,An improved model of the condensation nucleus counter with stereophotomicrographic recording. Geof. pura e appl.41 (1958), 211–216.
L. W. Pollak andA. L. Metnieks,The fog in the photoelectric nucleus counter, Geof. pura e appl.38 (1957), 181–203.
L. W. Pollak andA. L. Metnieks,New calibration of photoelectric nucleus counters, Geof. pure e appl.43 (1959), 285–301.
L. W. Pollak andA. L. Metnieks,Intrinsic calibration of the photoelectric condensation nucleus counter model 1957 with convergent light beam, Technical (Scientific) Note No. 9, Contract AF 61 (052)-26, Dublin Institute for Advanced Studies (1960).
L. W. Pollak andT. C. O'Connor,Photoelectric condensation nucleus counter of high precision, Geof. pura e appl.32 (1955), 139–146.
P. Squires,The growth of cloud drops by condensation, J., General Characteristics, Australian J. Sci. Res. [A] Phys. Sci.5, 1 (1952), 59–86.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Emmanuel, M.B., Squires, P. Absolute calibration of a photoelectric condensation nucleus counter at low values of nucleus concentration. PAGEOPH 72, 268–283 (1969). https://doi.org/10.1007/BF00875710
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00875710