1 Introduction

In the last two decades, laser diodes have become important in experimental science. Especially vertical-cavity surface-emitting lasers (VCSEL) have a versatile field of application. Compared to other laser diodes VCSELs have low production costs, high modulation efficiency and their circular beam profile allows simple fibre coupling [1]. Therefore, they are widely used in optical fibre communication, in spectroscopy [2] and in consumer electronics.

VCSELs are also interesting for space applications such as for data communications inside of satellites or new scientific instrument developments [3]. For this application, it has to be demonstrated that VCSELs can tolerate the harsh conditions in space and also provide the required lifetime of several years for space missions.

In this paper we summarize the results of a long-term endurance test under vacuum condition which was accomplished at the Optics- and Opto-Electronics Laboratory (OOEL) [4] of the European Science and Technology Centre (ESTEC) with VCSELs from three different manufacturers. The tested laser diodes were all single-mode VCSELs emitting at a wavelength of 795 nm hermetically sealed in a TO 46 package. The test was carried out under vacuum condition to enable testing as close as possible to the environmental conditions in space. For example, leakages in the sealing of the lasers could occur which would cause a loss of the package atmosphere and thus could have an influence on the lifetime [5].

This test was carried out in the context of the development of a new space magnetometer prototype, the Coupled Dark State Magnetometer (CDSM) [3, 6]. The CDSM is a type of coherently optically pumped magnetometer. Its functional principle is based on the measurement of the Zeeman energy level shift of the ground state within the Zeeman manifold of the 87Rb D1 hyperfine structure caused by an external magnetic field. This energy level shift is measured using a quantum interference effect called Coherent Population Trapping (CPT) [7, 8]. From this shift, the external magnetic field can be calculated by means of the Breit–Rabi formula [9].

A prototype has been developed by the Institute of Experimental Physics of the Graz University of Technology (TUG) and the Space Research Institute (IWF) of the Austrian Academy of Sciences.

The first demonstration of this new magnetometer in space will take place aboard the Chinese Seismo-Electromagnetic Satellite (CSES) mission in spring 2018.

The CDSM was also selected for the JUpiter ICy Moon Explorer (JUICE) mission of the European Space Agency (ESA) which will be launched in 2022. This mission will visit the Jovian system and will investigate, among others, the magnetic field around Jupiter’s moon Ganymede, which will help studying its inner structure including a presumed subsurface ocean.

For the development of the CDSM, VCSELs are used because their properties fit the requirements for the preparation of the CPT effect and also fulfil those for the use in space applications such as small size, low power consumption and robustness. For the excitation of the CPT effect we use a single-mode VCSEL emitting a carrier wavelength of λ = 794.979 nm (vacuum wavelength) [10]. On this carrier frequency we apply a modulation frequency of 3.417 GHz via the power supply current of the VCSEL, which corresponds to the half of the ground-state splitting frequency of 87Rb.

Therefore, one of the biggest advantages of the VCSEL for our application is its high modulation ability of several gigahertz. However, a disadvantage of this laser type for the use in CPT resonance excitation applications can be its typically higher laser line width of about 50–100 MHz compared to, e.g. DFB lasers (2–5 MHz). The laser line width should be less or equal to the homogenous line width of the excited optical resonance to allow an efficient preparation of the CPT effect [11, 12]. A further drawback is its high current tuning coefficient (see Table 5). Compared to edge emitters this coefficient is about a factor of ten higher. Thus, current noise stemming from the laser’s constant current source is more efficiently translated into frequency noise leading to a higher noise level in the magnetic field measurement.

For the JUICE mission it is important to prove that the VCSELs can operate for several years (up to 17 years) under vacuum condition and retain their optical properties required for a proper function of the CDSM. Therefore, the VCSELs were artificially aged under vacuum condition to investigate their lifetime and repeatedly electro-optically characterised in the course of the test [13].

In Sect. 2 a short summary of VCSEL’s reliability testing is given. Sections 3 and 4 summarize the test setup and the most important results.

2 VCSEL reliability and test parameters

In general, there are several reasons for failure of VCSEL diodes. Commonly, they are divided into failures which cause the diodes to malfunction suddenly and those that cause a gradual slow deterioration of the laser properties like optical power [14], which is called wear-out.

Abrupt failure is caused for example by growth of dislocation networks often described as dark line defects due to material defects, Electro Static Discharge (ESD) or mechanic damage [15, 16].

In general, the lifetime of a VCSEL significantly depends on its parameters of operation [17]. High junction temperatures and high current densities—especially in the active region—reduce the lifetime of a VCSEL. Thus, by increasing the VCSEL’s current and temperature compared to their typical operational values Iop and Top, the lasers aging can be accelerated and the VCSELs’ behaviour of several years of operation can be studied within a much shorter time. Depending on the chosen test parameters and the design of the VCSEL the lasers are aged at different rates. The factor by which the aging process is increased is called acceleration factor AF. It is defined as the ratio of the time-to-failure at nominal operational conditions TTFop to that at accelerated operational conditions TTFtest.

The acceleration factor reached for the VCSELs is calculated according to following formula [17]:

$${\text{AF}}=\frac{{{\text{TT}}{{\text{F}}_{{\text{op}}}}}}{{{\text{TT}}{{\text{F}}_{{\text{test}}}}}}={\left( {\frac{{{I_{{\text{test}}}}}}{{{I_{{\text{op}}}}}}} \right)^n}\exp \left( {\frac{{{E_{\text{A}}}}}{{{k_{\text{B}}}}}\left( {\frac{1}{{{\text{T}}{{\text{j}}_{{\text{op}}}}}} - \frac{1}{{{\text{T}}{{\text{j}}_{{\text{test}}}}}}} \right)} \right)$$
(1)

Here Itest is the applied laser current during the endurance test and Iop the typical (nominal) operational current value in our magnetometer application. The exponent n as well as the activation energy EA are empirically determined by the manufacturer during their reliability testing. kB is the Boltzmann constant, Tjop is the junction temperature for nominal operation and Tjtest the junction temperature for the test temperature. However, the junction temperature of the VCSEL cannot be measured directly and is calculated via the relation [1]:

$${\text{T}}{{\text{j}}_x}={T_x}+{R_{{\text{th}}}} \cdot \left( {{I_x} \cdot {V_x} - {P_x}} \right)$$
(2)

The junction temperature depends on the ambient temperature Tx and the heating of the laser chip by electrical power dissipation during its use. The heat that is generated by the laser itself is calculated with the thermal resistance Rth and the difference between the introduced electrical power (laser current Ix times laser voltage drop Vx) and the emitted optical power Px. For the calculation of the acceleration factor the parameters n, EA and Rth were provided by the VCSEL manufacturers [13].

The nominal operational parameters (Iop and Top) of each VCSEL were determined during the first electro-optical characterisation. The choice of the operational current was based on the required optical power Px for the operation of the magnetometer. The corresponding laser temperature Top was determined to achieve a wavelength of λ = 794.979 nm and varies for each laser which influences the obtained equivalent operational life time τeop (see Table 2).

The test was planned with four VCSELs at a time from three different suppliers. The abbreviations for the different suppliers which will be used further on and the corresponding part numbers as well as the operational currents Iop and the applied current during the endurance test Itest are stated in Table 1.

Table 1 Part numbers, operational laser currents Iop as well as laser current during the endurance test Itest for the used lasers

The test duration was first 1 year and then extended further for 2 months. The ambient temperature Tx during the vacuum test was set to 70 °C, which was the highest possible temperature of the test setup. The applied laser current Itest for VCSELs from supplier B and C was set to their maximum rating according to the data sheet. The reached equivalent operational life time τeop of each tested VCSEL and the mean acceleration factor AFmean can be found in Table 2.

3 Laser endurance test description

The main goals of the laser endurance test were to investigate the aging behaviour of VCSELs under vacuum condition and to determine if the optical and electrical properties of the laser degrade over the achieved lifetime. All used VCSELs are single-mode lasers emitting at 795 nm which are hermetically sealed in a TO 46 package.

Before the lifetime test 12 VCSELs in total (4 each from supplier A, B and C) were selected and electro-optically characterized. For the vacuum test, the selected lasers were mounted in two separate vacuum chambers at the Optics- and Opto-Electronics Laboratory (OOEL) of the European Science and Technology Center (ESTEC) [4]. Half of the lasers (2 from each supplier) were mounted in one vacuum chamber (see Fig. 1) where they remained permanently for the whole vacuum test duration, while the lasers in the second chamber were removed twice for electro-optical characterization.

Fig. 1
figure 1

Optical and electrical setup of the endurance test under vacuum condition: the lasers were mounted in mounting blocks on temperature-stabilized carriers. The mounting blocks contained lenses for the collimation of the laser beams and a Pt1000 element to measure the temperature. Once per hour the temperature Ttest, current Itest and voltage Utest of the lasers as well as the emitted optical power were measured. The polarization-independent Ptest,unpol and the polarization-dependent optical power Ptest,pol were detected outside the vacuum chamber by two photodiodes and one additional polarizer. Both photodiodes were moved between the windows by a linear translation stage

Two VCSELs each were assembled in mounting blocks which contained two biconvex lenses to form two collimated Gaussian laser beams and a Pt1000 element to measure the temperature of the mounting blocks individually. The mounting blocks were fixed inside the vacuum chamber on temperature-stabilized carriers. The mean temperature during the test was (65.6 ± 2.0) °C in vacuum chamber 1 and (66.1 ± 1.8) °C in vacuum chamber 2. Each VCSEL was supplied with its own custom-made current source providing the predefined currents (see Sect. 2).

The vacuum chambers have several optical windows which allowed the detection of the optical power outside of the chambers via photo diodes. One photo diode measured the total laser power while the other photo diode detected the laser polarization-dependent optical power with the help of an additional linear polarizer placed in front of the photo diode. Thus, changes in the orientation of the polarisation plane like the unwanted arbitrary polarization jumps could be measured. All VCSEL were mounted in such a way that the maximum of the linear polarization state was measured. For these measurements, the two optical detectors were moved from one optical window to the next by a linear translator.

Over the whole test duration the laser current Itest, laser voltage Utest, temperature Ttest, polarization-dependent laser power Ptest,pol and polarization-independent laser power Ptest,unpol were measured and logged by a data acquisition system once per hour. The recorded laser power of each laser as a function of test duration is presented in Sect. 4.1.

For all lasers, the selected electrical and optical parameters were measured before and after the accelerated aging test (preliminary and final electro-optical characterization). Additionally, VCSELs mounted in the vacuum chamber number 1 were characterized twice during the accelerated aging test [intermediate characterization (IC)]. These characterization measurements were performed outside the vacuum chamber on an optical bench. Therefore, the VCELSs were dismounted from the vacuum chamber mounting blocks and assembled into a laser setup which allowed changing and stabilizing both the laser temperature (within ± 1 mK) and laser current. The laser temperature was measured with a Pt1000 element using a four-wire sensing schema and was set with the help of a thermoelectric cooler driven by a temperature control unit (feedback loop). The laser current could be changed by an adjustable constant current source and was monitored with a precision multimeter to an accuracy of ± 10 nA.

The following measurements were performed during the electro-optical characterisation: laser current and temperature combinations for achieving a constant wavelength of λ = 794.979 nm (vacuum wavelength), current tuning coefficient, temperature tuning coefficient, laser line width, threshold current and polarization ellipse.

The determination of the emitted laser frequency was calibrated by measuring the resonant excitation inside a glass cell filled with isotopically pure 87Rb. To obtain resonant excitation of the rubidium atoms, a sawtooth-shaped modulation was applied onto the laser current to sweep the laser frequency. During the sweep the transmitted laser light through the cell was measured by a photo diode. The detected photo diode signal was depicted with an oscilloscope showing the 87Rb D1 spectrum. The 87Rb (52S1/2 F = 2 → 52P1/2 F = 1) transition was used as reference to set the laser frequency. At certain laser current values within the operational range of the laser, the laser temperature was adapted such that this selected transition was excited and the corresponding laser current and temperature values were read.

At the same measured laser current–temperature pairs, also the current tuning coefficient ΔνI and the temperature tuning coefficient ΔνT were determined. For the measurement of the current tuning coefficient the laser temperature was fixed and the difference of the laser current ΔI required to change the laser frequency from the 87Rb (52S1/2 F = 2 → 52P1/2 F = 1) transition to the 87Rb (52S1/2 F = 1 → 52P1/2 F = 2) transition measured. With the known frequency spacing (Δν = 7.650 GHz) between these two transitions, the current tuning coefficient ΔνI was calculated.

For the measurement of the temperature tuning coefficient ΔνT a similar approach was used. Now, the laser current was fixed and the difference in laser temperature required to change the laser frequency between these two transitions was measured. Again from the known frequency spacing Δν the temperature tuning coefficient ΔνT was calculated.

The laser line width was measured with the help of an optical spectrum analyser. The used device has a free spectral range of δFSR = 1980 MHz, finesse of F ≈ 300 and thus a resolution of approximately δν = 7 MHz. A measurement uncertainty of about ± 5 MHz could be obtained for the laser line width.

The threshold current was determined by making use of the appearance of the speckle pattern when increasing the laser current from zero to above the threshold current. If the laser is operated at a laser current below the threshold current, the laser diode emits virtually incoherent light. At the threshold current the laser starts emitting coherent light and a speckle pattern becomes visible on a rough surface. With the help of an infrared sensitive camera the appearance of the pattern was determined by slowly increasing the laser current from zero up to the threshold current. With this method the threshold current can be determined with an accuracy of ± 0.1 mA.

The polarization ellipse of the linear polarisation state of the laser light was also investigated. The maximum and the minimum of the ellipse were determined with the help of a Glan–Thompson polarizer and a power meter.

The obtained results of the electro-optical tests are presented in Sect. 4.2.

4 Test results

4.1 Long-term behaviour of optical laser power under vacuum condition and accelerated aging

In total the VCSELs in chamber 1 were operated for 344 days and the VCSELs in chamber 2 for 414 days. For the lasers in vacuum chamber 1, the first intermediate characterisation (1st IC) took place after 114 days and the second one (2nd IC) was carried out after 288 days. Chamber 2 was not opened during the entire test duration.

During the test, all lasers in both chambers were switched off for maintenance for 1 week each (chamber 1 after 114 and 275 days, chamber 2 after 147 and 308 days).

For the illustration of the measurement results in Fig. 2, the measured data were smoothed using a moving average over 20 readings because the raw data were noisy due to the applied measurement method with the linear translator (see Sect. 3). This artificial noise resulted from slight variations in the position of the detectors and was not caused by the VCSELs itself.

Fig. 2
figure 2

Behaviour of the optical power of all tested lasers during the accelerated endurance test under vacuum condition over the test duration Δttest: the left column contains the results from vacuum chamber 1 and the right column from vacuum chamber 2. Figures a and b show the long-term behaviour of the optical power of lasers from supplier A, c and d from supplier B and e and f from supplier C

The lasers were labelled with their respective supplier letter (A, B or C) and additionally the two lasers from each supplier which were mounted in vacuum chamber 1 got the numbers 1 and 2 and the two further lasers from the same supplier installed in vacuum chamber 2 received the numbers 3 and 4. Further, the label of the initially mounted lasers ends with a number 1 and the replaced lasers with a number 2. For example, laser number 2 from supplier A which was initially mounted in vacuum chamber 1 was named A2.1 and was replaced by laser A2.2. Figure 2 includes both the polarization-independent and polarization-dependent laser power, which is marked with an additional suffix p.

The laser failure criterion was defined as the reduction of the measured laser power below 50% of its initial value PI at Ttest. This power reduction can be tolerated by the setup and allows a proper operation of the magnetometer.

All lasers showed an initial reduction of their optical power when the laser temperature was increased to Ttest and the vacuum established in the chambers. Within the first few hours the laser power increased again and stabilized. As initial value the optical power after this transient behaviour was taken.

In the course of the intermediate characterisations the functioning lasers were optically and electrically measured and the failed lasers replaced.

The discontinuities of the optical power after the intermediate characterizations in Fig. 2 resulted from the complete dismounting of the lasers from the mounting blocks and a slightly different alignment after reassembly. Due to this discontinuities, the relative changes in optical power after the intermediate characterization were related to the new starting value after the intermediate characterization.

In Fig. 2a and b the results of the vacuum test for supplier A are summarized. Laser A1.1 showed only minor relative changes of its optical power (about 3%) during the test phases. In contrast the optical power of laser A2.1 started degrading immediately after mounting in the vacuum chamber and failed on the second day of operation in vacuum. Laser A2.1 was replaced during the 1st IC by laser A2.2. The optical power of laser A2.2 behaved quite stable until the last test month where its optical power reduced by about 5%.

Also the laser power of A3.1 only showed reduction of its optical power by about 3% until the 350th day of operation in the vacuum test. Then the laser power decreased faster until the 405th day (in total about 20%) and revealed a polarization jump. The polarization jump is visible as a leap in the laser polarization-dependent optical power and no change in the polarization-independent power. The polarisation of laser A4.1 revealed in total two polarisation jumps within the whole test period. The first polarization jump appeared shortly before the first switch off for maintenance after 144 days of operation. After switching on the lasers again the original polarization plane was restored. The second polarization jump appeared after 200 days and returned to its initial polarization plane on the 249th day (see Fig. 3). The change of the polarization state in Fig. 3 took up to 20 h as well as the return to the original polarization state. The change of the polarization state of laser A3.1 required as well 20 h. On the contrary, the first polarization jump of laser A4.1 took only 3 h.

Fig. 3
figure 3

Polarization jump detected in the run of the optical power of laser A4.1: the polarization-dependent power was reduced but the polarization-independent power was unchanged. After 49 days the original polarization state was restored

In total, laser A4.1 exhibited a gradual decrease of its optical power by about 9% by the end of the test.

Figure 2c and d show the long-term behaviour of the optical power under vacuum condition for supplier B. Laser B2.1 started degrading right from the test start and failed after 21 days of operation in the vacuum chamber. It was replaced by laser B2.2 which showed an unusual reduction of its optical power (− 16% of initial power PI = 142 µW) with a subsequent increase (+ 34%) until the 2nd IC. After the 2nd IC the optical power increased further by 20% (initial power after 2nd IC PI = 211 µW). Laser B1.1 gradually degraded within 174 and failed after the 1st IC. The replacement laser B1.2 was operated for 56 days under vacuum and showed a reduction of its optical power of 4%. The optical power of the laser B3.1 decreased over the whole test duration and reduced severely within the last test month. On the 403th day its output power reached values below the defined laser failure limit of 50% initial power and even reduced further to 60% initial power at the end of the test. Laser B4.1 also revealed a significant drop in laser power. The optical power dropped from its initial value of PI = 194 µW below the failure criterion after 254 days of operation and reached its minimum of 95 µW, which is just below the failure criterion (total reduction of 49.5%). Then the laser power increased again until the end of the test to 196 µW, which is even above the initial value.

Since lasers B4.1 and B2.2 showed a similar laser power variation as laser B3.1, it could be possible that also the optical power of laser B3.1 recovered again but as the test was not continued further, there is no way of telling.

In Fig. 2e and f the behaviour of the optical power of lasers from supplier C are outlined. All lasers of supplier C had in common that the optical power started to increase after the beginning of the endurance test (2–7% of PI) before it stabilized after several days. Therefore, an additional burn-in would help to improve the stability of the optical output power.

Laser C1.1 showed proper function until it suddenly failed after 40 days. The failed C1.1 laser was replaced during the 1st IC by laser C1.2 and was sent back to its manufacturer for further investigations.

The variations of the optical power for lasers C2.1 and C1.2 were within 1% of their initial power.

The noise of the polarization-independent optical power of lasers C3.1 and C4.1, which is not seen in the measurement of the polarization-dependent optical power, is assumed to be caused by a random misalignment of the laser beam with respect to the optical detector in connection with the repeatedly moved linear translation stage. First both lasers C3.1 and C4.1 showed the typical increase in optical power. Then the optical power of laser C3.1 gradually reduced to approximately its initial value and the optical power of laser C4.1 remained 3% above its initial value after the end of the test.

After the completion of the endurance test the equivalent operational lifetime τeop of the tested lasers was calculated with the mean acceleration factor AFmean and the test duration Δttest according to

$${\tau _{{\text{eop}}}}={\text{A}}{{\text{F}}_{{\text{mean}}}} \cdot \Delta {t_{{\text{test}}}}$$
(3)

For the calculation of the mean acceleration value AFmean, first the acceleration factor of each data point AF i according to Eq. (1) was calculated and then averaged.

In Table 2 the maximum equivalent operational life time τeop of each VCSEL is stated which would be reached based on the measurements at the end of this endurance test. Most of the lasers were still working within the defined pass/fail criterion (> 50% of initial output power PI). For those lasers which failed during the test, the maximum reached lifetime was calculated up to the time when they failed.

The maximum equivalent operational lifetime τeop in Table 2 mainly varies for each VCSEL due to the different operational temperatures Top required to reach the defined laser frequency at the defined operational current Iop. Thus, also the maximum equivalent operational lifetime is different for each laser.

Table 2 Calculated maximum equivalent operational lifetime τeop for all tested lasers at the end of the accelerated aging test: the factor AFmean is mainly influenced by the different operational temperatures Top of each laser

4.2 Test results from electro-optical characterisations

In this section the results from the electro-optical characterisations are discussed. No results for lasers A2.1, B2.1 and C1.1 are presented in this section, since they have only been characterized once due to their failure prior to the first intermediate characterisation point and, therefore, no information on their aging behaviour could be obtained. The lasers B3.1 and B4.1 showed a reduction of their optical power of more than 50% but were still working at the time of the electro-optical characterisation and, therefore, they were also tested. Additionally, lasers C3.1 and C4.1 are not mentioned in the discussion, since a mishandling of the lasers after the endurance test caused a failure of the lasers.

In Fig. 4 an example of the measured laser current and laser temperature combinations to set the wavelength of the 87Rb (52S1/2 F = 2 → 52P1/2 F = 1) transition as well as in the measurements of the current tuning coefficient are shown for laser A2.2. Both Fig. 4a and b contain the measurement results from the preliminary (τeop = 0 a), intermediate (τeop = 9.9 a) and the final characterisation (τeop = 13.1 a) of laser A2.2.

Fig. 4
figure 4

Laser A2.2: a measurements of the laser current and laser temperature combinations to set the defined wavelength, b measurements of the current tuning coefficient

In Fig. 4a the laser temperature Top of laser A2.2 showed in total a reduction of ΔTop = (− 3.1 ± 0.2) °C between the preliminary and the final characterisation. Also the current tuning coefficient changed for the aged laser by (24 ± 16)% (see Fig. 4b). Similar measurements have been carried out for all tested lasers. The change detected in the operational laser temperature Top at Iop and the current tuning coefficient are summarized in Table 5 in the Appendix. The change of Top at Iop for all lasers is additionally shown in Fig. 5.

Fig. 5
figure 5

Change of the operational laser temperature Top at Iop over the calculated equivalent operational lifetime: a lasers of supplier A, b lasers of supplier B and c lasers of supplier C

Since each laser experienced a different aging in the experiment, the change of Top over the equivalent operational lifetime τeop is depicted.

All lasers from supplier A (Fig. 5a) showed a reduction of Top over the test duration. The laser temperature Top of laser A3.1 shifted by (− 6.1 ± 0.2) °C. Lasers from supplier B showed no clear trend (Fig. 5b). The laser temperature Top of laser B4.1 as well as of laser B2.2 increased during the test but laser B1.1, which failed after the first intermediate characterisation, decreased.

Both tested lasers from supplier C showed an increase of Top.

5 Discussion

The lasers from the three suppliers performed differently during the endurance test and the electro-optical characterizations.

Most of the lasers from supplier A and supplier C showed a moderate decrease of their optical power during the lifetime test (< 10%). Lasers from supplier B revealed significant variations of their optical power (see Fig. 2). For example, the optical power from laser B4.1 decreased by 50% and increased again to approximately its initial value. Such strong variations in laser power could significantly influence the performance of the magnetometer.

Four lasers in vacuum chamber 1 completely failed (A2.1, B2.1, C1.1 and B1.1) and were replaced during the intermediate tests. Two lasers in vacuum chamber 2 (B3.1 and B4.4) showed a reduction in laser power more than 50% but were still working at the end of the endurance test and thus they were tested during the final electro-optical characterization. More lasers failed in vacuum chamber 1 than in vacuum chamber 2. However, all lasers in vacuum chamber 1 and 2 were handled in the same way and powered from identical equipment. There was no known difference between the two vacuum chambers.

Lasers A2.1 and B2.1 failed shortly after test start and laser B1.1 also showed a clear reduction of its optical power right from the start. Most likely these lasers were already damaged at the test start.

However, one laser (C1.1) suddenly failed after 40 days of operation in vacuum chamber 1. This laser was sent back to its manufacturer for further tests on the failure.

Investigations by the manufacturer indicated that this laser showed typical features of a laser suffered from an ESD event. Since the high sensitivity of VCSELs on ESD was known, several ESD precautions during handling and operation were made but due to the repeated handling and mounting of the lasers during the test, the risk of ESD damage cannot be eliminated completely. Therefore, the intermediate characterizations with the complete dismounting of the lasers entailed additional risks for ESD damage. Most of the lasers showed only a slight increase or reduction of their optical power after remounting which was explained by a different reassembly. However, a drop in the optical power, which for example laser A2.2 showed after the 2nd IC, could also be caused by a prior ESD event.

Due to the failure of the tested lasers, we decided to include measurements of the reverse current–voltage, forward current–voltage as well as the optical power–current characteristics in future test procedures. These additional measurements can help to detect defect or already degraded VCSELs before mounting them in a test setup. Especially, in the reverse current–voltage measurement a degradation, for example caused by an ESD event, can be identified by a flattening of the originally sharp breakdown voltage [17].

Two lasers from supplier A revealed polarization jumps (A3.1, A4.1). For applications which use polarization optics or a polarization-maintaining fibre, the preservation of the polarization plane is critical. In such applications polarization jumps could cause severe problems and result in a complete loss of the optical power. This was also one reason why in the CDSM application a multimode fibre, which connects the laser unit with the sensor unit, in combination with polarization optics at the sensor unit was chosen. After the multimode fibre the optical light is nearly unpolarised and the defined polarization state is obtained with the polarization optics in the sensor unit. Therefore, optical power is lost at the sensor unit but a failure caused by polarization jumps is avoided. However, the measurements have shown that the change of the polarization state could last up to 20 h. Further investigation would be required to study this transition area and to evaluate the influence on the operation of the magnetometer. Additionally, depending on the rate of polarization jumps they could influence the CPT resonance shape [18, 19].

With respect to the JUICE mission it was important to test if a VCSEL laser can operate up to 17 years under vacuum condition. The endurance test allowed us to prove that lasers from at least supplier A and C can reach lifetimes of more than 17 years and still have enough optical power for the operation of the CDSM magnetometer. Lasers A1.1, A3.1 and A4.1 from supplier A as well as lasers C3.1 and C4.1 from supplier C reached an equivalent operational lifetime of more than 17 years. The equivalent operational lifetime for lasers of supplier B was much lower than the lifetime of the two other suppliers A and C. This was due to the fact that the activation energy EA, which we had received from the manufacturer B, is about half the value of the two other manufacturers. Thus, the calculated equivalent operational lifetime is much lower for lasers of supplier B.

The measurement on the laser temperature Top at the operational laser current Iop for the required nominal wavelength of λ = 794.979 nm (vacuum wavelength) showed different trends for the three manufacturers (see Fig. 5). The laser temperature Top for supplier A decreased for all tested lasers over time. Lasers from supplier B showed both reduction and increase of the laser temperature Top. The tested lasers from supplier C revealed an increase of their laser temperature Top during the test campaign.

Laser A3.1 had the largest operational temperature shift of all tested lasers of (− 6.1 ± 0.2) °C for an equivalent operational lifetime of about 20 years. The reason for this large change could be caused by a mode jump, which this laser showed several times during the vacuum test. Nevertheless, in the CDSM application such a change of the operational temperature Top can be compensated. However, for different applications enough margins in the design of the instrument should be taken into account.

Also laser B4.1, which failed due to a reduction of its initial power below the defined criterion, showed a large increase of Top at Iop of (+ 5.6 ± 0.6) °C. However, laser B3.1, which also showed a power reduction of 50% initial power, has no change of its operational temperature. All other lasers showed far less changes of Top at Iop.

The current tuning coefficient ΔνI for the lasers of manufacturers A, B and C also revealed differences. In general, lasers from supplier C showed the smallest current tuning coefficient and were a factor 2–3 smaller than for suppliers A and B (see Table 5). A smaller current tuning coefficient is a benefit for the operation of the magnetometer because laser current noise is less efficiently converted into frequency noise.

Most of the lasers showed no change of their current tuning coefficient within the measurement accuracy after aging. However, the current tuning value ΔνI of lasers A3.1, A2.2, B1.1 and B3.1 increased (see Table 5; Fig. 4). The change of ΔνI, which is stated in column three of Table 3, was calculated between the initial characterization and the final characterization.

Table 3 Change of the current tuning coefficient ΔνI between initial and final electro-optical characterization: As was mentioned in Sect. 4.1 the lasers aged differently and thus they reached a different equivalent operational lifetime τeop at the end of the accelerated endurance test

In the CDSM application an increase in the current tuning coefficient would cause an increase of the noise level and thus in a reduction of the signal-to-noise ratio.

The mean temperature tuning coefficient ΔνT was identical for the lasers of the three suppliers (see Table 4). This parameter revealed a dependence neither on the laser current nor on the laser temperature and did not alter for increased operational lifetime.

Table 4 Mean temperature tuning coefficient ΔνT for the three different suppliers

The temperature tuning coefficient depends on the change of the average refractive index and the thermal expansion of the cavity with temperature [1]. The equality of the temperature tuning coefficients can possibly be explained by the fact that the material composition of the VCSELs is quite similar and thus they have almost the same linear temperature expansion coefficient. However, the current tuning coefficient seems to depend significantly on the exact inner structure of the lasers (e.g. size of the cavity, number of quantum wells, current confinement) and is thus unequal for the lasers from the three different suppliers.

VCSELs from supplier A showed typical threshold current values between 0.4 and 0.6 mA. From supplier C, the threshold current varied in the range from 1.3 to 1.5 mA and for supplier B from 0.7 to 0.8 mA. Lasers from supplier A as well as lasers C2.1 and C1.2 showed no change of their threshold current over time within the measurement accuracy. On the other hand, lasers B1.1, B3.1 and B4.1 from supplier B revealed a clear increase of their threshold current over their operational lifetime. The threshold current of VCSEL B1.1 and B3.1 increased from 0.8 to 1.3 mA and for laser B4.1 from 0.9 to 1.6 mA. An increase of the threshold current can either be a typical result of an aged laser [1] or by ESD-driven degradation [20]. Either way it is a hint that the laser structure is already degraded to a certain degree.

Lasers from suppliers A and B had a typical line width of about 100 MHz and VCSELs from supplier C about 50 MHz. A smaller laser line width is preferable for the CPT resonance excitation (see Sect. 2). The repeated measurement of the laser line width showed that the theoretical accuracy of ± 5 MHz was not reached. Due to the required readjustment of the optical path for this measurement, an experimental accuracy of about ± 10 MHz is obtained. Most of the laser line widths showed no significant change for aged lasers except for lasers B1.2, B2.2 and A4.1 which laser line widths’ seemed to narrow. The narrowing could also be caused by a back reflection from an optical element into the laser.

The polarization ellipse was evaluated by measuring its maximum and minimum. The ratio of the maximum to minimum was used to evaluate changes. However, the accuracy of the measurement of the minimum was influenced by background light and, therefore, here an accuracy of ± 2 µW was obtained. Within this limits no significant changes of the ratio were measured during the electro-optical characterisations.

Based on the results from the endurance test and the electro-optical characterizations we decided to further test lasers from supplier C for the CDSM prototype. Both lasers from suppliers A and C performed well during the endurance test. However, lasers from supplier C showed several benefits for the CDSM application compared to the other suppliers. They had no polarization jumps, less changes in the electro-optical parameters compared to the two other suppliers while aging, a lower current tuning coefficient, typically a two times smaller laser line width and a two to three times higher optical power than lasers from supplier A. The obtained output power from laser A was limited by the maximum allowed operational laser current (Imax = 2 mA) according to the data sheet.

An endurance test with a bigger lot of lasers from supplier C will be performed within a new laser vacuum test facility at Graz University of Technology. The benefit of this new test facility is that it allows measurements of the electro-optical characteristic without dismounting the lasers. Additionally, the test facility allows the performance of a further burn-in to stabilize the optical parameters.

6 Conclusion

In summary, 16 VCSEL were tested under accelerated operational conditions under vacuum. Half of the VCSELs were operated for 344 days with two intermediate electro-optical characterizations (vacuum chamber 1) and the other half for 414 days without interference (vacuum chamber 2).

The majority of VCSELs did function well until the end of the test. The optical output power of VCSELs from suppliers A and C behaved stable with slight changes during the vacuum test (< 10%) except laser A3.1 which showed a decrease of 20%. Lasers from supplier B showed the greatest variation in optical power during the accelerated aging test (up to 50%). Such strong variations could influence the performance of our magnetometer. Four lasers in vacuum chamber 1 failed completely (A2.1, B2.1, C1.1 and B1.1).

With the endurance test we were able to show that VCSELs from supplier C can survive more than 17 years of operation with no major degradation of their performance and thus they are well suited for long-term space missions like JUICE. However, a burn-in appeared to be needed for the selected supplier C to stabilize the optical power.

For the JUICE application we decided to further test lasers from supplier C and perform an endurance test with a bigger lot in a new laser vacuum test facility at Graz University of Technology. The benefit of this new test facility is that it allows measurements of the electro-optical characteristic in the endurance test setup without dismounting the lasers.