Abstract
The simplest type of insensitive gold-leaf electroscope, as used in radioactive work, is critically and experimentally investigated with a view to increasing the accuracy of measurements made with this instrument. Results are quoted, showing a consistency of the order of 1 in 1000; and highly accurate comparisons are shown to be possible between radioactive sources of widely differing magnitudes.