On the performance of large monolithic LaCl3(Ce) crystals coupled to pixelated silicon photosensors

, , , , , , , , and

Published 21 March 2018 © 2018 IOP Publishing Ltd and Sissa Medialab
, , Citation P. Olleros et al 2018 JINST 13 P03014 DOI 10.1088/1748-0221/13/03/P03014

1748-0221/13/03/P03014

Abstract

We investigate the performance of large area radiation detectors, with high energy- and spatial-resolution, intended for the development of a Total Energy Detector with gamma-ray imaging capability, so-called i-TED. This new development aims for an enhancement in detection sensitivity in time-of-flight neutron capture measurements, versus the commonly used C6D6 liquid scintillation total-energy detectors. In this work, we study in detail the impact of the readout photosensor on the energy response of large area (50×50 mm2) monolithic LaCl3(Ce) crystals, in particular when replacing a conventional mono-cathode photomultiplier tube by an 8×8 pixelated silicon photomultiplier. Using the largest commercially available monolithic SiPM array (25 cm2), with a pixel size of 6×6 mm2, we have measured an average energy resolution of 3.92% FWHM at 662 keV for crystal thicknesses of 10, 20 and 30 mm. The results are confronted with detailed Monte Carlo (MC) calculations, where optical processes and properties have been included for the reliable tracking of the scintillation photons. After the experimental validation of the MC model, we use our MC code to explore the impact of a smaller photosensor segmentation on the energy resolution. Our optical MC simulations predict only a marginal deterioration of the spectroscopic performance for pixels of 3×3 mm2.

Export citation and abstract BibTeX RIS

10.1088/1748-0221/13/03/P03014