Summary
A method for the determination of very low concentrations of U, Th, Fe, Zn, Tl, Cd, Cu and Ag in high purity aluminium with isotope dilution mass spectrometry (IDMS) is developed using a compact and cost-efficient thermal ionization quadrupole mass spectrometer. The detection limits obtained are (in ng/g):U=0.018, Th=0.06, Fe=82, Zn=86, Tl=0.2, Cd=4, Cu=1, Ag=2.6. By this method it is possible to determine the α-emitters U and Th in aluminium down to the sub-ng/g level with good precision of 0.4–10% and 0.5–5%, respectively. The results should also be accurate because IDMS is a reliable analytical method. The dissolution of aluminium is carried out by aqua regia followed by the trace/matrix separation and the isolation of the trace elements by anion exchange chromatography (U, Th, Zn, Tl, Cd), electrodeposition (Cu, Ag) and extraction (Fe). Different aluminium samples are analysed by IDMS and the results are compared with those of other methods.
Similar content being viewed by others
References
Ortner HM, Blödorn W, Friedbacher G, Grasserbauer M, Krivan V, Virag A, Wilhartitz P, Wünsch G (1988) Kontakte (Darmstadt) 38
Ortner HM, Wilhartitz P (1990) Fresenius J Anal Chem 337:686
Roeser K (1988) Aluminium 64:277
Genna JL (1988) J Crystal Growth 89:62
Burba P, Willmer PG (1985) Fresenius Z Anal Chem 322:266
Kudermann G (1988) Fresenius Z Anal Chem 331:697
Saisho H, Tanaka M, Nakamura K (1988) J Res Natl Bur Stand 93:398
Hutton RC, Eaton AN (1988) J Anal At Spectrosc 3:547
Takeda K, Yamaguchi T, Akiyama H, Masuda T (1991) Analyst 116:501
Egger KP, Krivan V (1987) Fresenius Z Anal Chem 327:119
Egger KP, Krivan V (1988) Fresenius Z Anal Chem 331:394
Honma N, Kurosawa S, Kawashima I (1982) Rev Elektr Commun Lab 30:503
Dumont J, Côté M, Hubert J (1989) Appl Spectrosc 43:1132
Zaray G, Burba P, Broekaert JAC, Leis F (1988) Spectrochim Acta 43B:255
Burba P, Willmer PG, Becker M, Klockenkämper R (1989) Spectrochim Acta 44B:525
Heumann KG (1988) In: Adams F, Gijbels R, van Grieken R (eds) Inorganic mass spectrometry. Wiley, New York, p 301
Heumann KG (1990) In: Günzler H, Borsdorf R, Fresenius W, Huber W, Kelker H, Lüderwald I, Tölg G, Wisser H (eds) Analytiker Taschenbuch, Vol. 9. Springer, Berlin Heidelberg New York, p 191
Heumann KG, Schindlmeier W, Zeininger H, Schmidt M (1985) Fresenius Z Anal Chem 320:457
Götz A, Heumann KG (1986) Fresenius Z Anal Chem 325:24
Götz A, Heumann KG (1988) Int J Mass Spectrom Ion Processes 83:319
Herzner P, Heumann KG (1992) Mikrochim Acta [Wien] 106:127
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Beer, B., Heumann, K.G. Trace analyses of U, Th and other heavy metals in high purity aluminium with isotope dilution mass spectrometry. Fresenius J Anal Chem 343, 741–745 (1992). https://doi.org/10.1007/BF00633556
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00633556