Brought to you by:
Paper The following article is Open access

Embedded 32-bit Differential Pulse Voltammetry (DPV) Technique for 3-electrode Cell Sensing

, , and

Published under licence by IOP Publishing Ltd
, , Citation Aqmar N.Z.N et al 2018 IOP Conf. Ser.: Mater. Sci. Eng. 340 012016 DOI 10.1088/1757-899X/340/1/012016

1757-899X/340/1/012016

Abstract

This paper addresses the development of differential pulse voltammetry (DPV) embedded algorithm using an ARM cortex processor with new developed potentiostat circuit design for in-situ 3-electrode cell sensing. This project is mainly to design a low cost potentiostat for the researchers in laboratories. It is required to develop an embedded algorithm for analytical technique to be used with the designed potentiostat. DPV is one of the most familiar pulse technique method used with 3-electrode cell sensing in chemical studies. Experiment was conducted on 10mM solution of Ferricyanide using the designed potentiostat and the developed DPV algorithm. As a result, the device can generate an excitation signal of DPV from 0.4V to 1.2V and produced a peaked voltammogram with relatively small error compared to the commercial potentiostat; which is only 6.25% difference in peak potential reading. The design of potentiostat device and its DPV algorithm is verified.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.