Laboratory Testing and Calibration of the Nuclei-Mode Aerosol Size SpectrometerThis grant was awarded to complete testing and calibration of a new instrument, the nuclei-mode aerosol size spectrometer (N-MASS), following its use in the WB-57F Aerosol Measurement (WAM) campaign in early 1998. The N-MASS measures the size distribution of particles in the 4-60 nm diameter range with 1-Hz response at typical free tropospheric conditions. Specific tasks to have been completed under the auspices of this award were: 1) to experimentally determine the instrumental sampling efficiency; 2) to determine the effects of varying temperatures and flows on N-MASS performance; and 3) to calibrate the N-MASS at typical flight conditions as operated in WAM. The work outlined above has been completed, and a journal manuscript based on this work and that describes the performance of the N-MASS is in preparation. Following a brief description of the principles of operation of the instrument, the major findings of this study are described.
Document ID
20000004424
Acquisition Source
Ames Research Center
Document Type
Other
Authors
Brock, Charles A.
(Denver Univ. Denver, CO United States)
Date Acquired
September 7, 2013
Publication Date
July 31, 1999
Subject Category
Instrumentation And Photography
Funding Number(s)
CONTRACT_GRANT: NAG2-1242
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.
No Preview Available