Iodine filter imaging system for subtraction angiography using synchrotron radiation

https://doi.org/10.1016/0168-9002(93)91245-IGet rights and content

Abstract

A new type of real-time imaging system was developed for transvenous coronary angiography. A combination of an iodine filter and a single energy broad-bandwidth X-ray produces two-energy images for the iodine K-edge subtraction technique. X-ray images are sequentially converted to visible images by an X-ray image intensifier. By synchronizing the timing of the movement of the iodine filter into and out of the X-ray beam, two output images of the image intensifier are focused side by side on the photoconductive layer of a camera tube by an oscillating mirror. Both images are read out by electron beam scanning of a 1050-scanning-line video camera within a camera frame time of 66.7 ms. One hundred ninety two pairs of iodine-filtered and non-iodine-filtered images are stored in the frame memory at a rate of 15 pairs/s. In vivo subtracted images of coronary arteries in dogs were obtained in the form of motion pictures.

References (19)

  • K. Umetani et al.

    Nucl. Instr. and Meth. A

    (1991)
  • W.-R. Dix et al.

    Nucl. Instr. and Meth. A

    (1992)
  • H.D. Zeman

    Nucl. Instr. and Meth.

    (1984)
  • R.K. Smither et al.

    Nucl. Instr. and Meth. A

    (1988)
  • W.M.J. Veigele

    Atomic Data Tables

    (1973)
  • E. Rubenstein

    SPIE 314 Digital Radiography

    (1981)
  • K. Umetani et al.

    Rev. Sci. Instr.

    (1992)
  • A.C. Thompson et al.

    SPIE 1140 X-ray Instrumentation

    (1989)
  • W. Thomlinson

    Rev. Sci. Instr.

    (1992)
There are more references available in the full text version of this article.

Cited by (0)

View full text